Alkaline phosphatase activity in pasteurized milk: a quantitative comparison of Fluorophos and colourimetric proceduresTools Payne, C. and Wilbey, R.A. (2009) Alkaline phosphatase activity in pasteurized milk: a quantitative comparison of Fluorophos and colourimetric procedures. International Journal of Dairy Technology, 62 (3). pp. 308-314. ISSN 1364-727X Full text not archived in this repository. To link to this article DOI: 10.1111/j.1471-0307.2009.00503.x Abstract/SummaryFluorophos and colourimetric procedures for alkaline phosphatase (ALP) testing were compared using milk with raw milk additions, purified bovine ALP additions and heat treatments. Repeatability was between 0.9% and 10.1% for Fluorophos, 3.5% and 46.1% for the Aschaffenburg and Mullen (A&M) procedure and 4.4% and 8.8% for the Scharer rapid test. Linearity (R-2) using raw milk addition was 0.96 between Fluorophos and the Scharer procedure. Between the Fluorophos and the A&M procedures, R-2 values were 0.98, 0.99 and 0.98 for raw milk additions, bovine ALP additions and heat treatments respectively. Fluorophos showed greater sensitivity and was both faster and simpler to perform.