Alkaline phosphatase activity in pasteurized milk: a quantitative comparison of Fluorophos and colourimetric proceduresPayne, C. and Wilbey, R.A. (2009) Alkaline phosphatase activity in pasteurized milk: a quantitative comparison of Fluorophos and colourimetric procedures. International Journal of Dairy Technology, 62 (3). pp. 308-314. ISSN 1364-727X Full text not archived in this repository. It is advisable to refer to the publisher's version if you intend to cite from this work. See Guidance on citing. To link to this item DOI: 10.1111/j.1471-0307.2009.00503.x Abstract/SummaryFluorophos and colourimetric procedures for alkaline phosphatase (ALP) testing were compared using milk with raw milk additions, purified bovine ALP additions and heat treatments. Repeatability was between 0.9% and 10.1% for Fluorophos, 3.5% and 46.1% for the Aschaffenburg and Mullen (A&M) procedure and 4.4% and 8.8% for the Scharer rapid test. Linearity (R-2) using raw milk addition was 0.96 between Fluorophos and the Scharer procedure. Between the Fluorophos and the A&M procedures, R-2 values were 0.98, 0.99 and 0.98 for raw milk additions, bovine ALP additions and heat treatments respectively. Fluorophos showed greater sensitivity and was both faster and simpler to perform.